Biosafety cabinet airflow calculator

Average inflow and downflow against the manufacturer’s nominal values (±5 fpm in the field), and downflow uniformity (each point within ±20% or ±16 fpm of the average, whichever is larger). The usual NSF/ANSI 49 field acceptance checks for a Class II cabinet.

Readings

Results

Inflow (average)100 ± 5 fpm102 fpmPASS
Downflow (average)65 ± 5 fpm65 fpmPASS
Downflow uniformityeach point within ±16 fpm of 65all 18 inPASS

The method behind it

Do this on every test without the calculator

Certorum checks each reading against NSF/ANSI 49, ASHRAE 110 or ISO 14644 as you type it on your phone, then issues the certificate and invoice when you sign off. Free for up to 40 units.