What a biosafety cabinet certification report should include
Updated · 3 min read
On this page
A certification report is what the lab hands its biosafety officer, its accreditor or an inspector to show that a cabinet protected people on a given day. If it's missing the instrument calibration dates or the nominal values the readings were judged against, it doesn't prove much, however good the readings were.
Here is what a complete Class II biosafety cabinet report contains, section by section.
1. The cabinet
- Manufacturer, model and serial number
- Class and type (A1, A2, B1, B2, C1)
- The customer's own asset or tag number
- Location: site, building and room
- How it's exhausted: recirculating, canopy (thimble) connection or hard-ducted
2. The standard and the criteria
- The standard and edition certified to (NSF/ANSI 49 and the year)
- The manufacturer's nominal inflow and downflow for this model, which the readings are judged against
- Any customer-specific criteria that differ from the defaults
3. Who certified it, and when
- Certifying company, with address and contact details
- Technician's name and NSF accreditation number (and its expiry)
- Date of the test
- Next due date
- Signature
4. Instruments
For each instrument used (thermal anemometer, DIM flow hood, aerosol photometer, aerosol generator, and any others): name or model, serial number and calibration due date. A reading taken with an instrument out of calibration doesn't count, so this table is the first thing an auditor checks.
5. The tests, each against its criterion
| Test | Record |
|---|---|
| Inflow | Method (DIM or the manufacturer's alternative), raw readings, any correction factor, the calculated velocity, the nominal and tolerance, pass/fail |
| Downflow | The grid of readings, the average against the nominal ±5 fpm, uniformity (±20% or ±16 fpm), pass/fail |
| Smoke patterns | Downflow, view-screen retention, work-opening edges and sash seals, each acceptable or not |
| HEPA filters | Supply and exhaust separately: aerosol, upstream concentration, scan or probe method, highest penetration found, pass/fail against 0.01% |
| Site installation | Exhaust alarm test results for canopy or ducted cabinets, and any placement concerns |
| Optional checks | Electrical, lighting, noise, vibration or UV if the customer asked for them |
The criteria behind these are explained in NSF/ANSI 49 field tests and acceptance criteria.
6. As found and as left
If you adjusted anything (blower speed, a damper, the sash stop), record the readings before and after. The as-found data answers the question the lab actually cares about: was this cabinet protecting people since the last visit?
7. The result, and what to do about a failure
- A pass/fail result for each test and an overall result
- For a failure: what failed, the likely cause, what was done, and whether the cabinet is safe to use until it's fixed
- Recommendations (a filter nearing the end of its life, a cabinet too close to a door)
8. The label on the cabinet
A label on the cabinet with the date, the result, the certifier and the next due date, so anyone using it can see at a glance. A QR code on the label that links to the report online saves the lab from digging through email when an inspector asks.
9. Corrections
If a report has to be corrected after it's issued (a typo in a serial, a missed reading), issue a revised report that says what changed and marks the earlier one as superseded, rather than quietly replacing the file. Auditors look for that trail.
Producing this without a template
Certorum builds this report from the readings you enter on site: cabinet details and nominal values from the unit's record, your accreditation number from your profile, instruments and their calibration dates from your instrument list, every test against its criterion, a QR code that checks the certificate online, and corrected reports that keep the original on file. It's free for up to 40 units.
Certorum does these checks as you type
Enter readings on a phone or tablet on site: each one is checked against the standard and the unit's own set points, and the certificate and invoice are ready when you sign off. Free for up to 40 units.
More guides
- ASHRAE 110 fume hood testing: procedure, ratings and acceptance criteria
- Biosafety cabinet classes and types explained
- How often do biosafety cabinets, fume hoods and cleanrooms need certification?
- Cleanroom air changes per hour: how to calculate and verify them
- ISO 14644-1:2015 cleanroom classification: class limits, sample locations and sample volume
- Biosafety cabinet certification: NSF/ANSI 49 field tests and acceptance criteria
- USP <797> cleanroom certification: what's tested every six months
- Handbook 44 scale test procedures
- NIST Handbook 44 scale tolerances explained
- Placed-in-service reports for scales
- Test weights for scale service