Biosafety cabinet classes and types explained
Updated · 3 min read
On this page
Before you can certify a biosafety cabinet you need to know exactly what it is, because the class and type decide the inflow it needs, how it's exhausted and which site installation checks apply. The type is on the cabinet's data plate; this guide explains what it means.
The three classes
| Class | Protects | How |
|---|---|---|
| Class I | The operator and the room | Room air is drawn in through the front opening, across the work and out through a HEPA filter. No product protection: unfiltered room air passes over the work. Inflow at least 75 fpm. |
| Class II | The operator, the product and the room | Inflow at the front opening plus HEPA-filtered downflow over the work surface; the exhaust is HEPA-filtered. The common laboratory and pharmacy cabinet. |
| Class III | Maximum containment | A gas-tight enclosure worked through attached gloves, under negative pressure, with HEPA-filtered supply and (often double) HEPA-filtered exhaust. Used for the highest-risk agents. |
NSF/ANSI 49 field certification is written for Class II cabinets, which is almost everything a field certifier sees.
Class II types
| Type | Minimum inflow | Air recirculated / exhausted | Connection | Chemical use |
|---|---|---|---|---|
| A1 | 75 fpm | About 70% recirculated, 30% exhausted | Exhausts into the room, or through a canopy (thimble) connection | Not for volatile chemicals |
| A2 | 100 fpm | About 70% recirculated, 30% exhausted | Into the room, or through a canopy connection | Minute amounts of volatile chemicals only when canopy-connected and approved by the lab's safety officer |
| B1 | 100 fpm | Most of the air (often 60–70%) exhausted, the rest recirculated | Hard-ducted to a dedicated building exhaust | Minute amounts, kept to the exhausted part of the work area |
| B2 | 100 fpm | None recirculated: total exhaust | Hard-ducted to a dedicated building exhaust | Where volatile chemicals are part of the work |
| C1 | 100 fpm | Part of the downflow exhausted through a dedicated area of the work surface, the rest recirculated | Usually a canopy-type connection; has its own exhaust fan | Per the manufacturer and the lab's assessment |
Type C1 was added to NSF/ANSI 49 in 2016. Because designs vary more than for the older types, follow the manufacturer's field test procedure and set points closely.
What the type changes when you certify
- Inflow criterion. The certificate checks inflow against the manufacturer's nominal value ±5 fpm, but a Type A1 can never go below 75 fpm and every other type below 100 fpm. A cabinet whose plate says A2 and whose nominal is 80 fpm is a red flag worth querying.
- Inflow method. Hard-ducted Type B cabinets often can't be measured with a direct inflow measurement (DIM) hood the same way as Type A; use the method the manufacturer specifies and record it.
- HEPA exhaust test. A Type A exhaust filter can usually be scanned from above; a hard-ducted Type B exhaust filter often can't, and is probe-tested in the duct instead.
- Site installation. A canopy connection needs its exhaust alarm checked and the gap confirmed to be drawing room air. A Type B needs the building exhaust proven and the exhaust alarm and blower interlock tested, because a Type B with no exhaust becomes an unprotected box blowing air out of its front opening.
The full test list and criteria are in biosafety cabinet certification (NSF/ANSI 49).
Record it on every unit
Certifying a "biosafety cabinet" without the class, type and nominal set points on record is how wrong criteria end up on a certificate. Certorum stores each cabinet's type and its manufacturer's nominal inflow and downflow on the unit, so every test is checked against that cabinet's own values.
Certorum does these checks as you type
Enter readings on a phone or tablet on site: each one is checked against the standard and the unit's own set points, and the certificate and invoice are ready when you sign off. Free for up to 40 units.
More guides
- ASHRAE 110 fume hood testing: procedure, ratings and acceptance criteria
- What a biosafety cabinet certification report should include
- How often do biosafety cabinets, fume hoods and cleanrooms need certification?
- Cleanroom air changes per hour: how to calculate and verify them
- ISO 14644-1:2015 cleanroom classification: class limits, sample locations and sample volume
- Biosafety cabinet certification: NSF/ANSI 49 field tests and acceptance criteria
- USP <797> cleanroom certification: what's tested every six months
- Handbook 44 scale test procedures
- NIST Handbook 44 scale tolerances explained
- Placed-in-service reports for scales
- Test weights for scale service