Cleanroom classification calculator

ISO 14644-1:2015 particle concentration limits for each class and size, the minimum number of sample locations for a room’s floor area, and air changes per hour from the airflow at each HEPA diffuser.

Room

Minimum sample locations (ISO 14644-1 Table A.1)6Air changes per hour33.8USP <797> asks for at least 30 ACPH in ISO 7 rooms and 20 in ISO 8.

Class limits, particles per m³

Class≥0.1 µm≥0.2 µm≥0.3 µm≥0.5 µm≥1 µm≥5 µm
ISO 11021———
ISO 2100241041—
ISO 31,000237102358—
ISO 410,0002,3701,02035283—
ISO 5100,00023,70010,2003,52083229
ISO 6—237,000102,00035,2008,320293
ISO 7———352,00083,2002,930
ISO 8———3,520,000832,00029,300
ISO 9———35,200,0008,320,000293,000

Limits are Cn = 10^N × (0.1/D)^2.08, rounded to three significant figures. Dashes are cells the standard leaves blank.