ISO 14644-1:2015 cleanroom classification: class limits, sample locations and sample volume
Updated · 3 min read
On this page
ISO 14644-1 classifies the air cleanliness of a cleanroom or clean zone by the concentration of airborne particles. The 2015 edition changed how sample locations are chosen and dropped the old 95% upper confidence limit calculation, so older procedures may need updating. This guide walks through a classification from start to finish.
Pick the occupancy state
A classification states which of three occupancy states it was done in:
- As-built: the installation is complete and running, with no equipment or people.
- At-rest: equipment installed and running as agreed, no people present.
- Operational: running normally, with the specified number of people working.
Pharmacy cleanrooms under USP <797> are certified under dynamic (operational) conditions. See USP <797> cleanroom certification.
Class limits
The maximum concentration for class N at particle size D (in µm) is:
Cn = 10^N × (0.1 / D)^2.08 particles per cubic metre,
rounded to no more than three significant figures. For the classes certifiers see most:
| Class | ≥0.3 µm | ≥0.5 µm | ≥1.0 µm |
|---|---|---|---|
| ISO 5 | 10,200 | 3,520 | 832 |
| ISO 6 | 102,000 | 35,200 | 8,320 |
| ISO 7 | 1,020,000 | 352,000 | 83,200 |
| ISO 8 | 10,200,000 | 3,520,000 | 832,000 |
Most pharmacy and lab work is classified at ≥0.5 µm. Classification at 5 µm is limited: low concentrations and particle losses in sample tubing make it unreliable, and the standard handles larger particles separately.
How many sample locations
The minimum number of locations depends only on the floor area of the room or zone (Table A.1 of the standard). Some points from it:
| Area up to (m²) | Locations |
|---|---|
| 2 | 1 |
| 4 | 2 |
| 6 | 3 |
| 8 | 4 |
| 10 | 5 |
| 24 | 6 |
| 52 | 10 |
| 76 | 15 |
| 104 | 16 |
| 156 | 20 |
| 352 | 24 |
| 1,000 | 27 |
Above 1,000 m², use 27 × (area ÷ 1,000), rounded up. Divide the room into that many sections of equal area and put one location in each, at a point representative of the section, at work height.
The cleanroom classification calculator gives the number of locations for a room's floor area (in m², with the ft² equivalent shown), alongside the class limits and air changes per hour.
How much air to sample
At each location, sample at least enough air to expect 20 particles at the class limit for the largest particle size you're classifying at:
Vs = (20 / Cn,m) × 1,000 litres, where Cn,m is the class limit at that particle size,
and never less than 2 litres or 1 minute. For ISO 7 at 0.5 µm that's 20 ÷ 352,000 × 1,000 ≈ 0.06 L, so the 2 L / 1 minute minimum governs; a 1 cfm (28.3 L/min) counter running for one minute is plenty. For ISO 5 at 0.5 µm it's about 5.7 L, still within a minute at 1 cfm.
Pass or fail
Each location is judged on its own: the concentration at each location (the average, if you took several samples there) must not exceed the class limit at every particle size you classified at. There's no longer a 95% upper confidence limit calculation across locations, as there was in the 1999 edition.
If one location fails, the room fails. Investigate (a leaking filter, a blocked return, people too close to the probe), fix it, and retest.
How often to reclassify
ISO 14644-2:2015 sets the maximum interval between classifications:
- ISO 5 and cleaner: 6 months
- ISO 6 to ISO 9: 12 months
The interval can be extended where the room is continuously or frequently monitored and stays within limits. Other tests (airflow, pressure difference, filter leakage, recovery) are scheduled by the owner's monitoring plan, and some industries are stricter: USP <797> requires every six months regardless of class.
Doing this on every job
Certorum works out the number of locations from each room's area, checks every count against the room's class limit as you enter it, flags the locations that are over, and puts the room back on its 6- or 12-month cycle when you sign off.
Certorum does these checks as you type
Enter readings on a phone or tablet on site: each one is checked against the standard and the unit's own set points, and the certificate and invoice are ready when you sign off. Free for up to 40 units.
More guides
- ASHRAE 110 fume hood testing: procedure, ratings and acceptance criteria
- Biosafety cabinet classes and types explained
- What a biosafety cabinet certification report should include
- How often do biosafety cabinets, fume hoods and cleanrooms need certification?
- Cleanroom air changes per hour: how to calculate and verify them
- Biosafety cabinet certification: NSF/ANSI 49 field tests and acceptance criteria
- USP <797> cleanroom certification: what's tested every six months
- Handbook 44 scale test procedures
- NIST Handbook 44 scale tolerances explained
- Placed-in-service reports for scales
- Test weights for scale service