USP <797> cleanroom certification: what's tested every six months
Updated · 4 min read
On this page
USP General Chapter <797>, Pharmaceutical Compounding — Sterile Preparations, sets the environmental requirements for pharmacies and other compounders. The revised chapter became official on November 1, 2023. Certification of the primary engineering controls (PECs) and the rooms around them is at the heart of it, and it's on a six-month cycle.
This guide summarizes what a <797> certification covers and the room requirements a certifier checks. It is a summary; the chapter, the compounder's own procedures and the state board of pharmacy have the final word.
What gets certified, and when
Primary engineering controls (PECs), the ISO Class 5 devices where compounding happens: laminar airflow workbenches, biological safety cabinets, compounding aseptic isolators (CAI) and compounding aseptic containment isolators (CACI).
Secondary engineering controls, the rooms they sit in: the ISO Class 7 buffer room, the ISO Class 8 (or 7) ante-room, and segregated compounding areas.
Certification is required:
- at least every 6 months,
- when a PEC or room is installed,
- when it is relocated or the room is redesigned,
- after major service, such as a HEPA filter change.
The tests
| Test | Applies to | What it shows |
|---|---|---|
| Airflow | PECs: velocity across the work zone. Rooms: air changes per hour from the HEPA supply | Enough clean air in the right direction |
| HEPA filter integrity | PEC filters and the room's ceiling HEPA filters | No leaks in the media or seals |
| Total airborne particle count | Each PEC and room, to ISO 14644-1, under dynamic operating conditions | The space meets its ISO class while people are working |
| Dynamic airflow smoke pattern test | Each PEC, under dynamic conditions | First air reaches the critical site; no turbulence or reflux into the work |
| Smoke at doors and pass-throughs | Room interfaces | Air moves from cleaner to less clean spaces |
| Pressure differential | Between adjacent rooms | Each room is pressurized the right way |
"Dynamic" means testing while compounding staff are present and working (or simulating it), not in an empty room. Particle counts in an at-rest room can pass while the same room fails with two people in it.
Room requirements
| Space | ISO class | Air changes per hour | Pressure |
|---|---|---|---|
| PEC | ISO 5 | Unidirectional airflow | — |
| Buffer room | ISO 7 | At least 30 total HEPA-filtered; at least 15 of them from the HVAC through ceiling HEPA filters | At least +0.020" w.c. to the ante-room |
| Ante-room | ISO 8 (ISO 7 if it opens into a negative-pressure buffer room) | At least 20 | At least +0.020" w.c. to unclassified areas |
The PEC's own HEPA-filtered air can count toward the buffer room's 30 air changes, but the HVAC must still supply at least 15 through the ceiling.
How to get the air change number right is in cleanroom air changes per hour. To check air changes from the diffuser readings, and the number of particle-count locations a room needs, use the cleanroom classification calculator. The method behind it is in ISO 14644-1 cleanroom classification.
Hazardous drugs: USP <800>
Rooms where hazardous drugs are compounded are kept negative so contamination stays in:
| Space | Requirement |
|---|---|
| Containment buffer room (C-SEC) | ISO 7, at least 30 ACPH, externally vented, between −0.010 and −0.030" w.c. relative to adjacent areas |
| Containment segregated compounding area (C-SCA) | Unclassified, at least 12 ACPH, externally vented, between −0.010 and −0.030" w.c. |
| Ante-room serving a C-SEC | ISO 7, positive to the unclassified areas around it |
| C-PEC | A Class II BSC or CACI, externally vented |
Viable sampling
Certification isn't the whole environmental program. <797> also requires viable air and surface sampling (air at least every six months for most compounders; surface at least monthly), often done by the same certifier. Results above these action levels need investigation and corrective action:
| ISO class | Air (CFU/m³) | Surface (CFU/plate) |
|---|---|---|
| ISO 5 | more than 1 | more than 3 |
| ISO 7 | more than 10 | more than 5 |
| ISO 8 | more than 100 | more than 50 |
The report
The compounder's inspector will look for, room by room and PEC by PEC: the ISO class achieved and the occupancy state tested, particle counts at each location, air changes, pressure differentials, HEPA results, smoke study results, the instruments and their calibration dates, and who certified it. Many certifiers follow CETA's application guides for sterile compounding facilities when writing these up.
Doing this on every job
Certorum keeps each pharmacy room and PEC on its six-month cycle, checks particle counts, air changes (from diffuser airflow or measured), pressure differentials and viable results against the room's class as you enter them, and produces a site summary report covering every room and PEC from the visit. Viable samples sent to a lab are noted on the report as results to follow, and the room stays on your list until the results are added on a revised report.
Certorum does these checks as you type
Enter readings on a phone or tablet on site: each one is checked against the standard and the unit's own set points, and the certificate and invoice are ready when you sign off. Free for up to 40 units.
More guides
- ASHRAE 110 fume hood testing: procedure, ratings and acceptance criteria
- Biosafety cabinet classes and types explained
- What a biosafety cabinet certification report should include
- How often do biosafety cabinets, fume hoods and cleanrooms need certification?
- Cleanroom air changes per hour: how to calculate and verify them
- ISO 14644-1:2015 cleanroom classification: class limits, sample locations and sample volume
- Biosafety cabinet certification: NSF/ANSI 49 field tests and acceptance criteria
- Handbook 44 scale test procedures
- NIST Handbook 44 scale tolerances explained
- Placed-in-service reports for scales
- Test weights for scale service