ASHRAE 110 fume hood testing: procedure, ratings and acceptance criteria
Updated · 4 min read
On this page
ANSI/ASHRAE Standard 110, Method of Testing Performance of Laboratory Fume Hoods (current edition 2016), is the test most chemical hygiene plans and lab ventilation programs point to. It is a method, not a pass/fail standard: it tells you how to measure a hood's containment, and the acceptance criteria come from the owner's specification, their chemical hygiene plan or ANSI/ASSP Z9.5, Laboratory Ventilation.
This guide covers the tests, the rating format and the criteria you'll most often be asked to meet. Always test to the criteria in the customer's own program.
As manufactured, as installed, as used
The same test is run under three conditions, and the result is labelled with the one used:
- AM — as manufactured: in the manufacturer's test facility, before shipping, under controlled conditions.
- AI — as installed: in the lab, connected to the building exhaust, with the room as designed but no one working at the hood.
- AU — as used: in the lab as it is normally run, with the hood loaded with the user's equipment and the room's normal activity.
A hood that passes AM can still fail AI if the lab's supply air blows across its face, which is why owners commission hoods in place.
The tests
1. Face velocity
The sash opening (at the design sash position) is divided into a grid of equal-area cells, roughly a square foot each, and the air velocity is measured at the center of each cell with a thermal anemometer, averaged over time at each point. The report gives the average and the spread across the opening. Uneven readings point to obstructions, a baffle problem or cross-drafts.
The fume hood face velocity calculator works out the average and the spread from your grid.
2. Flow visualization
- Local smoke: a small smoke source is passed along the sash edges, the airfoil, the work surface and the walls. Smoke should be drawn back into the hood without lingering, rolling forward or escaping.
- Large-volume smoke: a smoke generator inside the hood fills it, showing whether the exhaust clears it evenly and whether any reaches the face.
3. Tracer gas containment
The core of ASHRAE 110. Sulfur hexafluoride (SF6) is released at 4.0 L/min from an ejector placed inside the hood, 6 inches behind the sash plane, at the left, center and right of the opening. A mannequin stands in front of the hood with a detector probe at its breathing zone, and the tracer concentration is logged at each position (typically for 5 minutes). A probe is also run around the perimeter of the opening to find local escape.
4. Sash movement effect
With the ejector at the center, the sash is opened and closed and the detector records the spike in concentration. A hood with a slow exhaust response (a variable air volume hood whose controls lag, for example) shows it here.
The rating
Results are written as the condition, the release rate and the control level. For example, AI 4.0 0.08 means as installed, tested at 4.0 L/min, with a control level of 0.08 ppm (the highest average concentration at the breathing zone).
Usual acceptance criteria
| Test | Criterion most programs use |
|---|---|
| Face velocity | Set by the owner. Commonly an average of 80–120 fpm at the design sash height (often 100 fpm), sometimes with each point within ±20% of the average |
| Smoke visualization | No visible escape at the face and no reverse flow out of the hood |
| Tracer gas | 0.05 ppm or less as manufactured; 0.10 ppm or less as installed or as used, at 4.0 L/min |
| Sash movement | Set by the owner; record the peak and the time it takes to recover |
| Airflow monitor | Reads correctly against your measurement and alarms on low flow |
Routine checks vs. a full ASHRAE 110
A full tracer-gas test needs SF6, a detector and a mannequin, so it's usually done at commissioning, after a hood or the lab's HVAC is modified, and on whatever periodic schedule the owner's program sets. The routine annual check most labs ask for is lighter:
- face velocity at the design sash height,
- smoke visualization at the face,
- the airflow monitor and its alarm,
- a sash-height sticker showing where the hood was tested.
Record what you actually did. A report that says "ASHRAE 110" for a face-velocity-only check sets the customer up for trouble in an audit.
How often
Most lab programs test fume hoods every 12 months, plus after installation, modification or a change to the building's exhaust. See how often each kind of unit needs certification.
Doing this on every job
Certorum checks the face velocity readings against each hood's own minimum and maximum at its design sash height as you enter them, records the smoke and monitor checks, and issues the report when you sign off.
Certorum does these checks as you type
Enter readings on a phone or tablet on site: each one is checked against the standard and the unit's own set points, and the certificate and invoice are ready when you sign off. Free for up to 40 units.
More guides
- Biosafety cabinet classes and types explained
- What a biosafety cabinet certification report should include
- How often do biosafety cabinets, fume hoods and cleanrooms need certification?
- Cleanroom air changes per hour: how to calculate and verify them
- ISO 14644-1:2015 cleanroom classification: class limits, sample locations and sample volume
- Biosafety cabinet certification: NSF/ANSI 49 field tests and acceptance criteria
- USP <797> cleanroom certification: what's tested every six months
- Handbook 44 scale test procedures
- NIST Handbook 44 scale tolerances explained
- Placed-in-service reports for scales
- Test weights for scale service