Test weights for scale service
Updated · 2 min read
On this page
Every scale test is only as good as the weights on the platform. Handbook 44 judges the scale; the weights have rules of their own. This guide covers what a registered service agency needs to have, how calibration works, and what to keep on record.
As with everything in weights and measures, your state's rules are what count. The references here are the national models: NIST Handbook 44 and Handbook 130 (2026 editions).
What weights you need
The model registration regulation in NIST Handbook 130 asks an applicant to have sufficient field standards and equipment to test the devices they service, as set out in the notes of the relevant Handbook 44 code. That equipment should meet the NIST 105-series specifications. For field standard weights, that's NIST Handbook 105-1, which covers field standard weights (Class F).
In practice, that means enough certified weight to run the increasing-load, decreasing-load and shift tests on the scales you service. For large scales, it's supplemented by substitution or strain-load testing (see Handbook 44 scale test procedures).
Calibration and traceability
Under the model regulation (Section 9):
- every field standard you use must be submitted to the weights and measures director for initial and subsequent verification and calibration, at intervals the director sets;
- you must not use field standards or test equipment that haven't been calibrated or verified;
- instead of taking the weights in, the director may accept calibration reports from a formally accredited or recognized laboratory, and keeps a list of the labs it accepts;
- the state can still check your standards when they're first put into service or at any point between calibrations.
Calibration intervals therefore vary by state, and sometimes by the type of standard. Know yours, and plan around it: a lapsed weight kit means no testing until it's recalibrated.
The one-third rule
Handbook 44's Fundamental Considerations (Appendix A, 3.2) recommend keeping standards accurate enough that corrections aren't needed. When a standard is used without a correction, its combined error and uncertainty must be less than one-third of the applicable device tolerance. When a correction is applied, the uncertainty of the corrected value must be under one-third of the tolerance.
This matters most on fine scales. The acceptance tolerance on a Class III scale with e = 0.01 lb is ±0.005 lb at small loads, so the weights used there need to be good to well under 0.0017 lb.
What to keep on record
- each weight or kit's identification, nominal value and class;
- the calibration certificate: who calibrated it, the date, the results, and the date it's due again;
- which weights you used on each test, so a test can be traced back to the certificate if anyone asks;
- any damage, cleaning or adjustment between calibrations.
Looking after them
- Keep weights clean, dry and covered; dirt, corrosion and damage all change mass.
- Handle them with gloves or lifting gear, not bare hands, especially smaller weights.
- Don't use a weight that's been dropped or damaged until it has been checked.
The scale tolerance calculator shows the tolerance at each test load, which tells you how good your weights need to be.
Certorum for scale service agencies
Handbook 44 tolerances checked as you enter each test load, as-found and as-left results on one report, placed-in-service reports, test weights on record and the invoice from the same visit. Free for up to 40 scales.
More guides
- Handbook 44 scale test procedures
- NIST Handbook 44 scale tolerances explained
- Placed-in-service reports for scales
- ASHRAE 110 fume hood testing: procedure, ratings and acceptance criteria
- Biosafety cabinet classes and types explained
- What a biosafety cabinet certification report should include
- How often do biosafety cabinets, fume hoods and cleanrooms need certification?
- Cleanroom air changes per hour: how to calculate and verify them
- ISO 14644-1:2015 cleanroom classification: class limits, sample locations and sample volume
- Biosafety cabinet certification: NSF/ANSI 49 field tests and acceptance criteria
- USP <797> cleanroom certification: what's tested every six months