Placed-in-service reports for scales
Updated · 3 min read
Most states let a registered serviceperson or service agency put a commercial scale into use, or back into use after a weights and measures official rejected it, without waiting for an official inspection. The price of that privilege is paperwork: a placed-in-service report sent to the state, so the official knows the device exists and can inspect it.
The model rule comes from NIST Handbook 130, Uniform Laws and Regulations in the Areas of Legal Metrology, in the "Uniform Regulation for the Voluntary Registration of Servicepersons and Service Agencies for Commercial Weighing and Measuring Devices". States adopt it with their own changes, so always check your state's version; the details below are the model's (2026 edition).
Who can place a scale in service
Under the model regulation, a holder of a Certificate of Registration may:
- remove an official rejection tag or mark from a device;
- place in service, until an official examination can be made, a device that had been officially rejected;
- place in service a new or used device that has just been installed.
The registrant is responsible for installing, repairing and adjusting devices "as closely as practicable to zero error". Registration can be suspended or revoked for, among other things, failing to have test standards certified, using inadequate test equipment, or failing to bring a device into compliance with Handbook 44 after service.
When to file
A report is required for each rejected device restored to service and for each newly installed device placed in service.
In the model rule:
- the report includes the registration number and is signed by the registered serviceperson (or a serviceperson representing a registered agency);
- within 24 hours, the original goes to the weights and measures director, together with any official rejection tag removed from the device;
- a copy goes to the device's owner or operator, and the servicer keeps a copy.
Some states take the report online, or have changed the deadline, the number of copies or what has to be attached. The 24-hour clock is the model; your state's rule is what counts.
What the report records
States design their own forms, but they generally ask for:
- who and when: service agency and technician, registration numbers, date placed in service, signature;
- where: device owner, business name and address, location of the device;
- what: device type, make, model, serial number, capacity, e and accuracy class, and, where marked, the NTEP Certificate of Conformance number;
- why: newly installed, or restored after rejection (with the rejection tag or its number);
- the test: that it was tested to Handbook 44 and found within tolerance, often with the test loads and results.
If you keep the test results anyway, the report is mostly a copy of what you already know. The slow part is retyping it onto a state form, which is exactly what software should do for you.
Good practice
- Fill it in on site, from the test you just ran, not from memory at the end of the week.
- Keep your copy with the test record and the weights you used, so an inspector or a customer can see the whole story.
- Track devices you've placed in service, so you can tell an official exactly what you did when they come to inspect.
The test itself is covered in Handbook 44 scale test procedures, and the weights you test with in test weights for scale service.
Certorum for scale service agencies
Handbook 44 tolerances checked as you enter each test load, as-found and as-left results on one report, placed-in-service reports, test weights on record and the invoice from the same visit. Free for up to 40 scales.
More guides
- Handbook 44 scale test procedures
- NIST Handbook 44 scale tolerances explained
- Test weights for scale service
- ASHRAE 110 fume hood testing: procedure, ratings and acceptance criteria
- Biosafety cabinet classes and types explained
- What a biosafety cabinet certification report should include
- How often do biosafety cabinets, fume hoods and cleanrooms need certification?
- Cleanroom air changes per hour: how to calculate and verify them
- ISO 14644-1:2015 cleanroom classification: class limits, sample locations and sample volume
- Biosafety cabinet certification: NSF/ANSI 49 field tests and acceptance criteria
- USP <797> cleanroom certification: what's tested every six months